Fluke Australia Pty Ltd—Fluke 33X Clamp Meters
PRA number: 2009/10778
Date created: 4
th May 2009
Product information
Product description
Fluke 333, 334, 335, 336 and 337 Digital Clamp Meters manufactured between 3 March 2008 and 27 February 2009.
Identifying features
Fluke 333, 334, 335, 336 and 337 Digital Clamp Meters
What are the hazards?
Electrocution due to incorrect reading
What are the defects?
A defect has the potential, over time, to cause the meter to display a low or no voltage reading on a circuit energised with a hazardous voltage. In this situation it is imperative that the practice of testing the meter on a known live circuit before and after checking the circuit under test is maintained. Should the meter not pass the "known live circuit test" its use should be discontinued immediately, identified as faulty and removed from service.
Where the product was sold
Supplier
Fluke Australia Pty LtdSupplier's web site
http://www.fluke.com.auWhat should consumers do?
Cease using and immediately return suspect products to the place of purchase to be replaced free of charge. For further information visit www.fluke.com.au or phone (02) 8850 3333.